9 - 11
IM 4H3B1-01E
9. CHANGING INITIAL SETTINGS
9.2.3 How to Set Internal or External RJC
The purpose of the following setting is to select whether internal or external reference
junction compensation values are used.
In the case of INTernal, the reference junction compensation is based on the internal
RJC circuit.
In the case of EXTernal, the reference junction compensation is based on an external
voltage input.
The initial value is ‘INT’.
SETTING PROCEDURE:
Enter the SET UP Mode by turning ‘ON’ the power while pressing and holding the
ENT-key until the SET UP display appears.
Use the UP/DOWN-keys to select the display ‘SETUP=RJC’. Press the ENT-key.
Select the channel No. for which you want to specify this RJC setting. Then press the
ENT-key.
Select INT or EXT using the UP/DOWN-keys. Press the ENT-key.
If you selected INT:
The RJC setting has been completed, but has not been stored yet. The display
‘*RJC SET*’ will appear. You can now adjust other settings in the SET UP Mode, by
pressing the ESC-key.
Before leaving the SET UP Mode, you have to store your new settings. Press the ESC-
key to return to the ‘SETUP=RJC’ display and then select the ‘SETUP=END
display.
Press the ENT-key. Select ‘END=STORE’ to keep your new settings or
END=ABORT’ and press the ENT-key. After a few seconds, the Operation Mode will
appear.
If you selected EXT:
Set the input voltage to use for reference junction using the UP/DOWN-keys (must be
within –20000µV to +20000µV range, initial value is 0µV). Then press the ENT-key.
The RJC setting has been completed, but has not been stored yet. The display
‘*RJC SET*’ will appear.You can now adjust other settings in the SET UP Mode, by
pressing the ESC-key.
Before leaving the SET UP Mode, you have to store your new settings. Press the ESC-
key to return to the ‘SETUP=RJC’ display and then select the ‘SETUP=END
display.
Press the ENT-key. Select ‘END=STORE’ to keep your new settings or
END=ABORT’ and press the ENT-key. After a few seconds, the Operation Mode will
appear.
9.2 How to Change Initial Settings for the Input

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA µR1800 and is the answer not in the manual?

YOKOGAWA µR1800 Specifications

General IconGeneral
BrandYOKOGAWA
ModelµR1800
CategoryData Loggers
LanguageEnglish

Related product manuals