11 - 2 IM 4H3B1-01E
11.2 Fuse Replacement
Replace the fuse at least once every two years for preventive maintenance. To replace
the fuse, turn OFF the main power supply.
1 Remove the chart compartment.
Do this as follows:
While gently pressing the chart paper compartment lock tab, located in the
middle at the lower front, in the direction of the arrow, swing and lift the
compartment from the unit.
2 The fuse is now visible in the right lower front of the plate. See also figure 11.2.
Turn the knob of the fuse holder counterclockwise, and the holder will slide out with the
fuse.
3 Make sure that the new fuse rating is correct and mount the new fuse by turning the
knob clockwise.
Figure 11.2 Fuse Replacement
Before replacing the fuse, make sure to turn OFF the power supply and
disconnect the power source. Use only specified fuses which should be
obtained from your nearest Sales & Service Office. The use of other fuses might
cause fire.
!
WARNING