Page #102 background imageLoading...
Page #102 background image
9 - 22 IM 4H3B1-01E
9.8 How to Change Initial Settings for the REMOTE Option
This paragraph describes the way to change initial settings for the REMOTE option.
The settings concerning the REMOTE option are always accessible, and data entry will
be ignored if your µR1800 is not equipped with this option.
The /R1 option allows you to execute functions on the µR1800 by a remote trigger
(contact input).
The functions which can be executed by a remote trigger are:
start/stop the recording (see also 9.8)
change the chart speed to a preset second chart speed
have messages printed out (up to 5 messages)
get a manual printout
get a periodic printout (see also 9.4.6)
start/stop TLOG computation (only in combination with the MATH option)
trigger the IC memory card (only in combination with the IC memory card option).
However, the /R1 terminal has only five terminals, which means that only five functions
can be triggered. The printing out of one message takes up one terminal.
Initially the functions are assigned to the terminals as follows:
1: start/stop the recording (RECORD)
2: change the chart speed to a preset second chart speed (CHART_SPD)
3: get a manual printout (MANUAL_PR)
4: have message 1 printed out (MESSAGE1)
5: have message 2 printed out (MESSAGE2)
The purpose of this setting is to assign the to be triggered functions to the five terminals.
SETTING PROCEDURE:
Enter the SET UP Mode by turning ‘ON’ the power while pressing and holding the
ENT-key until the SET UP display appears.
Use the UP/DOWN-keys to select the display ‘SETUP=REM’. Press the ENT-key.
Select the terminal No. (CONT 1 to 5) ( /R1 has five terminals.) Press the ENT-key.
Select the function which you want to assign to that terminal. You can choose from
RECORD (start/stop recording) (see also 9.9), CHART_SPD (second chart speed),
MESSAGE1, 2, 3, 4 or 5 (message No. 1, 2, 3, 4, 5), MANUAL_PR (manual printout),
DGT_PR (periodic printout), TLOG or IC_TRIG.
Press the ENT-key. The display ‘*REM SET*’ will appear.
This setting has been completed, but has not been stored yet. Press the ENT-key to set
other terminals or press the ESC-key to adjust other settings in the SET UP Mode.
Before leaving the SET UP Mode, you have to store your new settings. Press the ESC-
key to return to the ‘SETUP=REM’ display and then select the ‘SETUP=END
display.
Press the ENT-key. Select ‘END=STORE’ to keep your new settings or
END=ABORT’ and press the ENT-key. After a few seconds, the Operation Mode will
appear.