Code Message Chapter or Section
850 Cannot be set or executed at current store state.
To set or execute, reset store.
16.4
852 Stored file is illegal.
Initialize memory before storing.
16.4
854 Waveform display data not found. —
855 Data destination memory is full.
Saving has been stopped.
—
856 An error has occured while storing.
Storing has been stopped.
—
857 Cannot be set while Master/Slave Synchronized
Measurement is set to Slave.
7.6
858 Store process is in progress now.
Execute or set setting again.
—
859 Cannot convert selected file.
Select a file with an extension of WTS or HDS.
16.3
862 Numeric data not found. —
863 Cannot be set or executed when different types of elements are installed. —
864 This wiring cannot be set as the first selected element. 1.1
865 Cannot be set while integration is running.
Stop or reset Integration.
8.3
866 Cannot be set or executed while Auto Print is operating.
Turn off Auto Print from the [PRINT MENU] (SHIFT+PRINT) menu.
19.2
867 Auto Print is not in operation.
Start Auto Print from the [PRINT MENU] (SHIFT+PRINT).
19.2
868 Print out destination must be set to Built-in Printer in order to start Auto Print.
Set [Print to] to Built-in from the [PRINT MENU] (SHIFT+PRINT) menu.
19.2
869 Auto Print funciton is not supported in the current measurement mode or settings. 19.2
870 Auto Print [Interval] setting is invalid.
Set [Interval] time to an appropriate amount from the [PRINT MENU](SHIFT+PRINT)
->[Auto Print Settings] menu.
19.2
871 Attempted made to start Auto Print when the end time had already passed.
Set [End Time] to a future date & time from the [PRINT MENU](SHIFT+PRINT)
->[Auto Print Settings] menu.
19.2
872 Auto print’s print-out has been canceled.
The printer or file system is in action.
19.2
874 Sync source, PLL source or trigger source cannot be set to Ext Clk, while Master/Slave
Synchronization Measurement is set to Slave.
7.6
875 Master/Slave Synchronization Measurement cannot be set to Slave, while sync
source, PLL source or trigger source is set to Ext Clk.
7.6
876 Can not calculate from present point value. 3.1 or 4.1
877 Can not set 0 to count. 15.1 or 16.1
879 Can not set or execute while recording high speed data.
Stop measurement and wait for file status "Ready".
15.4
880 Cannot be set or executed while initialization.
Wait until status changes to "Ready".
15.4
881 Cannot be set or executed while measurement is in progress.
To set or execute, "Stop" measurement.
15.4
882 Stopped measurement. Detection error of measuring interval signal.
Check External Sync (MEAS START) input.
4.4*
883 Cannot be set or executed in High Speed Data Capturing Mode. —
884 Can not set wiring to 1P3W/3P3W in High Speed Data Capturing Mode.
Select a different wiring.
1.1
885 Cannot be set or executed in High Speed Data Capturing Mode.
Set or execute in Normal Measurement Mode.
Appendix 10*
886 Cannot be set or executed to same current ranges, due to different types of elements
are installed or external current sensor settings are not same.
1.3
887 Cannot start integration. Turn off Independent Element setting by the [WIRING] menu,
or switch the measurement ranges to fixed ranges.
1.1
888 Cannot start the integration. Turn off Independent control by the [INTEG] menu or turn
off Auto of the Data Update Inerval by the [UPDATE RATE] menu.
8.1 or 1.15
889 Setting and execution is not available when auto-ranging is set to ON. 1.2