EasyManua.ls Logo

ADVANTEST R3162

Default Icon
588 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
R3132 Series Spectrum Analyzer Operation Manual
3.3.26 SWEEP Key (Sweep Time)
3-75
Gate Position Sets the position of a gate signal.
Gate Width Sets the width of a gate signal.
Gated Sweep ON/OFF Turns the gated sweep mode on or off.
ON: Sweep is executed according to the currently set gate
conditions (gate position and width).
OFF: Turns the gated sweep mode off.
Screen A/B Toggles the active screen between A and B.
A: The upper screen is set to the active state.
B: The lower screen is set to the active state.
Gate Setup Quit The trigger condition setup screen of gate signals is released to re-
turn to the Sweep menu.
Gated Sweep ON/OFF Turns the gated sweep mode on or off.
ON: Sweep is executed according to the currently set gate
conditions (gate position and width).
OFF: Turns the gated sweep mode off.
Ext Gate In ON/OFF Changes the input signal at the External Trigger connector.
ON: Performs a gated sweep using the External Trigger con-
nector signal as the gate signal.
Sweeps intermittently when the signal level is greater
than approximately 2.5V, and stops sweeping when the
signal level is approximately less than 2.5V.
OFF: Performs a normal sweep.
Uses the input signal at the External Trigger connector
as the trigger signal for the Ext Trig mode.

Table of Contents

Related product manuals