R3132 Series Spectrum Analyzer Operation Manual
5.4.2 Linearity Error (Internal Mixer Mode)
5-113
Table 5-31 Relationship Between the Frequency Range and Signal Generator Frequency Setting
(Used on the R3132 Series)
Range
Signal generator
1st Frequency
Signal generator
2nd Frequency
1 kHz/ 29.996 MHz 30.004 MHz
2.5 kHz/ 29.99 MHz 30.01 MHz
5 kHz/ 29.98 MHz 30.02 MHz
10 kHz/ 29.96 MHz 30.04 MHz
25 kHz/ 29.9 MHz 30.1 MHz
50 kHz/ 29.8 MHz 30.2 MHz
100 kHz/ 29.6 MHz 30.4 MHz
250 kHz/ 29 MHz 31 MHz