R3132 Series Spectrum Analyzer Operation Manual
5.2.10 Scale Fidelity
5-40
Table 5-16 Linear Scale Fidelity Test Setting
div. from Reference Level Signal Level (nominal)
00 dB
1 -0.92 dB
2 -1.94 dB
3 -3.10 dB
4 -4.44 dB
5 -6.02 dB
6 -7.96 dB
7 -10.46 dB
8 -13.98 dB
9 -20.00 dB