EasyManuals Logo

ADVANTEST R3162 User Manual

Default Icon
588 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #377 background imageLoading...
Page #377 background image
R3132 Series Spectrum Analyzer Operation Manual
5.2.16 Frequency Response
5-66
61. After the auto tuning has completed, adjust signal generator output level so that
the marker reading is -10 dBm ±0.09 dB.
62. Record the display of power meter reading with reverse sign in the performance
verification record sheet.
63. On the R3172/82, press FREQ and to increase center frequency by 200 MHz.
64. On the signal generator, increase the frequency of output by 200 MHz.
65. On the power meter, increase the correction frequency by 200 MHz.
66. Repeat steps 60 through 65 for every center frequency by 200 MHz step up the
center frequency to 26.4 GHz (for R3172) or 26.8 GHz (for R3182) listed in the
performance verification record sheet.
Measuring frequency response in the frequency range: 27 GHz to 30 GHz for R3182
(Note: For testing this frequency range, preselector tune is required)
67. On the R3182, set center frequency to 27 GHz.
68. On the signal generator, set the frequency to 27 GHz.
69. On the power meter, set the correction frequency to 27 GHz.
70. On the R3182, press as follows to tune the preselector.
FREQ, 1/2_more, Presel, Auto Tune
71. After the auto tuning has completed, adjust signal generator output level so that
the marker reading is -10 dBm ±0.09 dB.
72. Record the display of power meter reading with reverse sign in the performance
verification record sheet.
73. On the R3182, press FREQ and to increase center frequency by 200 MHz.
74. On the signal generator, increase the frequency of output by 200 MHz.
75. On the power meter, increase the correction frequency by 200 MHz.
76. Repeat steps 70 through 75 for every center frequency by 200 MHz step up the
center frequency to 29.8 GHz listed in the performance verification record sheet.
Measuring frequency response in the frequency range: 30 GHz to 40 GHz for R3182
(Note: For testing this frequency range, preselector tune is required)
77. On the R3182, set center frequency to 30 GHz.
78. On the signal generator, set the frequency to 30 GHz.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ADVANTEST R3162 and is the answer not in the manual?

ADVANTEST R3162 Specifications

General IconGeneral
BrandADVANTEST
ModelR3162
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals