R3132 Series Spectrum Analyzer Operation Manual
5.2.19 Gain Compression
5-80
8. On the signal generator (SMP02), turn output level on.
9. On the signal generator (SMP02), adjust output level until the signal level at 2.5
division in the left hand part on the R3132 series screen is decreased by 1 dB from
-30 dBm.
10. Remove the RF cable from the input terminal of R3132 series, connect power
sensor there.
11. On the power meter, set the correction frequency to 200 MHz.
12. Record the level of the power meter reading on the performance verification
record sheet.
Under preamplifier on condition
Setting the measurement conditions (Frequency 200 MHz)
13. Connect the signal generators as shown Figure 5-19.
14. On the both of signal generators, Set controls as follows:
Signal Generator (SMP02)
Frequency: 201 MHz
Output Level: -50 dBm
Signal Generator (SMP04)
Frequency: 200 MHz
Output Level: -50 dBm
15. On the R3132 series, after preset, set controls as follows:
Center Frequency: 200.5 MHz
Span: 2 MHz
Reference Level: -50 dBm
Hi Sense: ON
Input Attenuator: 0 dB
dB/div: 1 dB/div
Measuring the gain compression (Frequency 200 MHz)
16. On the signal generator (SMP02), turn output level off.
17. On the signal generator (SMP04), adjust the output level for a displayed signal of
-50 dBm ±0.1 dB on the R3132 series screen
18. On the signal generator (SMP02), turn output level on.
19. On the signal generator (SMP02), adjust output level until the signal level at 2.5
division in the left hand part on the R3132 series screen is decreased by 1 dB from
-50 dBm.