R3132 Series Spectrum Analyzer Operation Manual
5.2.19 Gain Compression
5-82
31. On the signal generator (SMP02), adjust output level until the signal level at 2.5
division in the left hand part on the R3162/72/82 screen is decreased by 1 dB
from -30 dBm.
32. Remove the RF cable from the input terminal of R3162/72/82, connect power
sensor there.
33. On the power meter, set the correction frequency to 3600 MHz.
34. Record the level of the power meter reading on the performance verification
record sheet.