EasyManua.ls Logo

ADVANTEST R3162

Default Icon
588 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
R3132 Series Spectrum Analyzer Operation Manual
5.3.2 Output Level Flatness
5-93
4. On the R3132 series, set controls as follows.
Center Frequency: 30 MHz
Span Zero: Span
TG Output Level: -10 dBm
TG: ON
5. On the power meter, set the correction frequency to 30 MHz.
6. On the power meter, set relative measurement mode.
Measuring the Tracking Generator Output Level Flatness
7. On the R3132 series, set center frequency to 100 kHz.
8. On the power meter, set the correction frequency to 100 kHz.
9. Record power meter reading on the performance verification record sheet.
10. Repeat steps 7 through 9 for each center frequency listed in Table 5-27.

Table of Contents

Related product manuals