EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>agilent 7683B

Agilent Technologies agilent 7683B User Manual

Agilent Technologies agilent 7683B
202 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #122 background imageLoading...
Page #122 background image
122 Installation, Operation, and Maintenance
Part 2, Operation
6 Vials and Bottles
How Many Sample Vials Can I Run?
The number of sample vials that you can run at one time is determined by the
following parameters for your application:
The number of sample injections per sample vial.
The syringe size used: 1 µL, 5 µL, 10 µL, 25 µL, 50 µL, or 100 µL.
The syringe wash volume (see Table 18)
The number of solvent washes (both pre- and post-injection) required from
each solvent bottle.
The number of sample washes and solvent washes per sample injection that
the injector discards into each waste bottle.
Number of waste bottles.
Table 1 8 Syringe wash volumes
Syringe Size µL 80% syringe
volume µL
40% syringe
volume µL
30% syringe
volume µL
20% syringe
volume µL
10.80.40.30.2
5421.51
10 8 4 3 2
25 20 10 7.5 5
50 40 20 15 10
100 80 40 30 20

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies agilent 7683B and is the answer not in the manual?

Agilent Technologies agilent 7683B Specifications

General IconGeneral
BrandAgilent Technologies
Modelagilent 7683B
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals