122 Installation, Operation, and Maintenance
Part 2, Operation
6 Vials and Bottles
How Many Sample Vials Can I Run?
The number of sample vials that you can run at one time is determined by the
following parameters for your application:
The number of sample injections per sample vial.
The syringe size used: 1 µL, 5 µL, 10 µL, 25 µL, 50 µL, or 100 µL.
The syringe wash volume (see Table 18)
The number of solvent washes (both pre- and post-injection) required from
each solvent bottle.
The number of sample washes and solvent washes per sample injection that
the injector discards into each waste bottle.
Number of waste bottles.
Table 1 8 Syringe wash volumes
Syringe Size µL 80% syringe
volume µL
40% syringe
volume µL
30% syringe
volume µL
20% syringe
volume µL
10.80.40.30.2
5421.51
10 8 4 3 2
25 20 10 7.5 5
50 40 20 15 10
100 80 40 30 20