144 Installation, Operation, and Maintenance
Part 3, Maintenance and Troubleshooting
8 Maintenance
Removing a Syringe
1 Loosen the plunger screw and raise the plunger carrier loop off the syringe
plunger.
2 Open the syringe latch.
3 Carefully lift the syringe flange out of the flange guide until clear, then lift
the syringe needle out of the needle support guide.
CAUTION
Be careful not to bend the syringe needle. Only pull the syringe out of the carriage until
clear. The needle bends easily when still seated in the needle support guide.