EasyManua.ls Logo

Agilent Technologies agilent 7683B - Cool On-Column Injection

Agilent Technologies agilent 7683B
202 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Special Topics 11
Installation, Operation, and Maintenance 183
Part 3, Maintenance and Troubleshooting
Cool On-Column Injection
For GCs with a cool on-column inlet, the 7683A/B injector can perform
injections directly onto 250-µm, 320-µm, and 530-µm columns. For a list of
consumables needed to perform these injections, see “Replacing Syringe
Needles" on page 107.
When performing cool on-column injections, the injector:
Slows the carriage speed so the overall injection time increases to
500 milliseconds.
Lowers the tip of the syringe needle an additional 19 mm into the column.
If you need to adapt your injector for cool on-column injection, change the
inlet column diameter, or change the septum nut or cool on-column insert,
refer to “Adapting For Cool On-Column Injection" on page 139.

Table of Contents

Related product manuals