96 Installation, Operation, and Maintenance
Part 2, Operation
4 Controlling the Sampler
2 The default value, 0 mm, includes a small safety factor to avoid striking the
bottom of the vial. This parameter can also be used to sample headspace
instead of the liquid or solid sample.
3 Enter your choices, then press Esc to return to the previous screen.
Dwell times
1 Display this screen.
Status / Automation / Injector / Dwell Times
2 Enter your choices. For most uses, both dwell times will be zero. This gives
a fast injection with a minimum of boil-out from a hot needle.
3 Press Esc to return to the previous screen.
To configure the injector
Depending on the GC firmware you have, the injector type and firmware, and
the turret size of your injector, you can set various solvent parameters. To
access these parameter settings: