Controlling the Sampler 4
Installation, Operation, and Maintenance 95
Part 2, Operation
Wash parameters
1 Display this screen.
Status / Automation / Injector / Washes
2 The syringe can be washed with sample before drawing up the amount to be
injected. It can also be washed with solvent both before (pre-washes) and
after (post-washes) an injection. The order of events is:
Wash syringe with solvent A Solvent-A-Pre-Washes times
Wash syringe with solvent B Solvent-B-Pre-Washes times
Rinse syringe with sample Sample-Pre-Washes times
Draw up sample and make injection
Wash syringe with solvent A Solvent-A-Post-Washes times
Wash syringe with solvent B Solvent-B-Post-Washes times
3 Enter your choices, then press Esc to return to the previous screen.
Needle depth (Sampling depth offset)
1 Display this screen.
Status / Automation / Injector / Depth Offset