Procedure 7. Installing the G2615A Bar Code Reader
1 Locate the recess in the bottom front of the tray. The BCR must be located
at this location. See Figure 23.
2 Locate the mounting screw holes. Some G2614A trays have two or three
holes that are flush with the bottom of the recess. Those trays require the
spacer. Newer G2614A trays have raised areas around three mounting holes
and do not need the spacer. The G2916A tray when upgraded for use with a
6890 GC requires a spacer.
The tray shown in Figure 23 is the two hole version. This style tray requires
use of the spacer to mount the BCR.
3 If the spacer is needed, align the holes in the spacer with the mounting
holes in the bottom of the tray. Place the BCR mounting bracket under the
spacer, insert the two long screws and tighten them firmly. See Figure 24.
If the spacer is not needed, place the BCR mounting bracket under the
mounting holes in the bottom of the tray. Insert the three short screws and
tighten them firmly. See Figure 24.
Figure 23 The BCR spacer and mounting location