Installation 2
Installation, Operation, and Maintenance 35
Part 1, Installation
Procedure 2. Installing the G2916A Tray
5890 Series II GC
When the sample tray is installed, it blocks access to the pneumatics door on
the left side panel. To maintain easy access to these components, Agilent
recommends the following:
1 Remove the GC left side panel.
2 Remove the pneumatic compartment access door from the left side panel.
3 Reinstall the left side panel.
4 Install the tray post in the hole in the right front corner (as you face the
oven) of the bracket (Figure 8).
5 Set the tray on the injector bracket so that the frontmost pin on the injector
bracket fits into the mating hole in the bottom of the tray, then align the
tray to the second alignment pin on the injector bracket.