124 Installation, Operation, and Maintenance
Part 2, Operation
6 Vials and Bottles
N
SS
= Total number of pre- and post-solvent A or solvent B washes and
sample washes per sample injected. When using both solvent A and B, the
sample wash is divided evenly between all four waste bottles in the
1-sample turret
Example
A tray is installed with a 1-sample turret in the injector. Your application
parameters are:
Two injections per vial
Three washes from solvent bottle A
Two washes from solvent bottle B
Two sample washes
10-µL syringe
80% wash volume (default)
1 Substitute the parameters of your application into the solvent bottle
equation:
For Solvent A
V
W
= 8 from Table 18
N
SI
= 2
N
SW
= 3
Maximum number of sample vials per solvent bottle
For Solvent B
V
W
= 8 from Table 18
N
SI
= 2
N
SW
= 2
Maximum number of sample vials per solvent bottle
2000
82× 3×
---------------------=41=
2000
82× 2×
---------------------=62=