EasyManuals Logo

Agilent Technologies agilent 7683B User Manual

Agilent Technologies agilent 7683B
202 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #125 background imageLoading...
Page #125 background image
Vials and Bottles 6
Installation, Operation, and Maintenance 125
Part 2, Operation
2 Substitute the parameters of your application into the waste bottle
equation:
For Waste A
V
Waste
= Use 8000 for the 1-sample turret
V
Wash
= 8 from Table 18
N
SI
= 2
N
SS
= 3 (solvent A washes) + 1 (Sample wash) = 4
Maximum number of sample vials per waste bottles used (WA)
For Waste B
V
Waste
= Use 8000 for the 1-sample turret
V
Wash
= 8 from Table 12
N
SI
= 2
N
SS
= 2 (solvent B washes) + 1 (sample wash) = 3
Maximum number of sample vials per waste bottles used (WB)
3 Analyze the results keeping the following in mind:
The 1-sample turret has 2 waste A bottles and 2 waste B bottles. If solvent A is
used, both waste A bottles must be used. If solvent B is used, both waste B bottles
must be used.
Sample washes are evenly distributed to all waste bottles used.
The injector can be configured to use 1, 2, or 3 bottles of solvent A or solvent B or
both.
Tables 19 and 20 summarize the results.
8000
82× 4×
---------------------=125=
8000
82× 3×
---------------------=166=

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies agilent 7683B and is the answer not in the manual?

Agilent Technologies agilent 7683B Specifications

General IconGeneral
BrandAgilent Technologies
Modelagilent 7683B
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals