68 Installation, Operation, and Maintenance
Part 1, Installation
2 Installation
Making A Trial Run
Once installation, configuration and updating are complete, make a quick
injection using the sampler to verify that it works properly.
1 Install an empty syringe in the injector.
2 If using a tray with a G2913A injector, install the 1-sample turret.
If using a tray with a G2613A injector, install the 3-sample turret.
3 Place empty bottles in all the Solvent A and Waste A turret positions. Place
an empty capped sample vial in the tray 1 position (or the turret sample 1
position, if not using the tray).
4 Make the sampler settings shown in Table 11. These are designed for a
6890N GC. If you are using a different GC, use these settings as a guide. To
enter the settings:
5890 Series II GC Use the G2911AA software.
For 6850 GC Use the handheld controller.
For 6890 GCs Use the GC keyboard.
Table 1 1 Trial run parameters
Parameter Setting
Injection volume 1
#Sample pumps 1
Viscosity delay 0
#Sample washes 1
#Solv A washes 1
#Solv B washes 0
Slow plunger Off
Pre-dwell time 0
Post-dwell time 0
Sampling offset Off
#Solv A pre-wash 1
#Solv B pre-wash 0
#Injections/vial 1
Samples 1–1
(for 6890 GC with tray and Agilent ChemStation
control, enter 1 in vial box and run method).