Vials and Bottles 6
Installation, Operation, and Maintenance 123
Part 2, Operation
Solvent bottle equation (G2913 A injector)
This equation estimates the maximum number of sample vials you can run
from one solvent bottle.
Maximum number of sample vials per solvent bottle
where:
V
W
= Volume of Wash in µL taken from Table 18
N
SI
= Number of sample injections taken from each sample vial
N
SW
= Number of Solvent washes (pre and post) per sample injected
Waste bottle equation
This equation estimates the maximum number of sample vials you can run
from one waste type.
Maximum number of sample vials per waste bottles used (WA or WB)
where:
V
Waste
= Volume of waste bottles used (WA or WB) in µL. Use 4000 for the
8-sample turret and 8000 for the 1-sample turret
V
Wash
= Volume of wash in µL taken from Table 18
N
SI
= Number of sample injections taken from each sample vial
CAUTION
The number of sample vials given by these equations are estimates. Solvent
characteristics such as evaporation rate and surface tension may affect the capacity of
the bottles.
2000
V
W
N
SI
× N
SW
×
----------------------------------------=
V
Waste
V
Wash
N
SI
× N
SS
×
----------------------------------------------=

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies agilent 7683B and is the answer not in the manual?

Agilent Technologies agilent 7683B Specifications

General IconGeneral
BrandAgilent Technologies
Modelagilent 7683B
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals