Faults and Errors 9
Installation, Operation, and Maintenance 161
Part 3, Maintenance and Troubleshooting
Fault light is on. The injector door is open.
Turret is not recognized if door
is closed.
1 Ensure that the injector door is
closed.
2 Re-install turret.
3 If the fault light stays on, obtain
Agilent service.
Fault and Run lights
are on.
The injector is mounted
incorrectly on the mounting
post.
Incorrect mounting post.
1 Ensure that the injector is
mounted properly. For more
information, see “Procedure
3. Installing the Injectors” on
page 42.
2 Ensure the correct mounting post
is installed. See “Procedure
3. Installing the Injectors” on
page 42.
3 If the Fault and Run lights stay on,
obtain Agilent service.
Align Mode light is on. The system was not initialized.
There is an injector memory
error.
Customer pushed the align
mode button.
1 Verify the turret is properly
installed. See “Changing the
Turret” on page 147.
2 Perform the alignment procedure
to initialize the system. See
Align Mode” on page 150.
3 If alignment fails, obtain Agilent
service.
All lights are on There is a board failure.
There is a firmware revision
conflict.
1 Check all cable connections.
2 Turn the instrument off, then on
again.
3 If the lights remain on, obtain
Agilent service.
Table 2 4 Status lights (continued)
Lights Possible cause Action