180 Installation, Operation, and Maintenance
Part 3, Maintenance and Troubleshooting
11 Special Topics
Use Figure 74 to determine the water bath temperature.
CAUTION
If the sample temperature is critical for accuracy, experiments should be performed to
validate water bath settings. Actual heat transfer varies with vial material, humidity,
condensation on the vial, and flow rates.
Figure 74 Vial temperature versus water bath and ambient air temperatures
Desired vial
temperature,
°C
Water bath temperature, °C
Ambient air
temperature,
°C
50
40
30
20
10
0
50403020100
40
30
25
20
15
10