86 Installation, Operation, and Maintenance
Part 2, Operation
4 Controlling the Sampler
After setting required parameters, select Send Settings to ALS to load the values
into the G2912A Controller. For more information, see the online Help.
If the controller somehow becomes corrupted with bad parameters, it may be
necessary to reset the controller to its factory defaults. To restore the G2912
Controller factory defaults, turn the controller power off, press and hold the
[Start] and [Stop] buttons, turn the power on, hold the buttons in for five
seconds and release the buttons.
See “Using the ALS Controller" on page 131 for information on using the
[Start] and [Stop] buttons on the controller and 5890GC.
Figure 37 5890 ALS controller software main screen