2 Setting up the Bravo Platform
Creating or adding a Bravo device
22 Bravo Platform User Guide
Set the desired values for the following properties. Use the default values for the
remaining properties.
Do not change the default value of the Teachpoint for robot Bravo property. For
example, the deck location 1 teachpoint identity is 1, the identity of deck location 2 is 2,
and so forth.
Property Description
Allowed/prohibited
labware
Optional. Click if you want to specify labware
restrictions for this location. The Allowed/prohibited
labware dialog box appears. For details on the labware
classes, see the VWorks Automation Control Setup Guide.
Note: If the button is not visible, click the empty field.
BCR on side If a barcode reader is set up at this location, you specify
the side.
Stack Height Type the maximum height (mm) for a labware stack at
this location.
Teachpoint for robot
<robot name>
If the device is integrated with an external robot, specify
whether the robot can access this location.
In Bravo Diagnostics, specify whether the Bravo head
should move to a safe location or can perform
operations at other deck locations concurrently.
Carefully plan which locations the external robot can
access.