2-2
387366-389 Operator Manual Installation Verification
2.2 Technical
Specifications
2.2.1 Electrical Service Electric single-phase service to the unit must be as specified on
the Equipment Drawing and on the Machine Data Plate.
Electric single-phase service requires a clearly marked
disconnect with lockout/tagout capability located near the
sterilizer.
Electric single-phase service should be on a separate circuit,
and not tied into circuits containing large reactive loads
(e.g., motors).
Sterilizer protective ground must be connected to terminal block
TB-1 in the sterilizer power box. Use green/yellow wire for
European installations.
3-phase service requires a clearly marked disconnect with
lockout/tagout capability located near the sterilizer.
2.2.2 Sterilizer Final Check Chamber must be leveled properly.
Door must open and close smoothly.
Door locked switches must be adjusted correctly.
Chamber strainer must be in place.
Rack and shelves and/or loading car operates correctly.
Warranty labels properly applied.
2.2.3 Cycle Operation Unit powers up correctly.
Run Leak Test cycle–leak rate is to be less than 1.0 mm
Hg/minute (1.3 mbar/minute). Only on prevac sterilizers.
Verify operation of a typical cycle (121° C[250°F] gravity).
2.2.4 Overall Size and Weight
(W x H x D)
16 x 16 x 26" Sterilizer: 660 x 1892 x 902 mm
(26 x 74.5 x 35.5")
20 x 20 x 38" Sterilizer: 762 x 1892 x 1152 mm
(30 x 74.5 x 45.375")