2-2
387366-389 Operator Manual Installation Verification
2.2 Technical
Specifications
2.2.1 Electrical Service • Electric single-phase service to the unit must be as specified on
the Equipment Drawing and on the Machine Data Plate.
• Electric single-phase service requires a clearly marked
disconnect with lockout/tagout capability located near the
sterilizer.
• Electric single-phase service should be on a separate circuit,
and not tied into circuits containing large reactive loads
(e.g., motors).
• Sterilizer protective ground must be connected to terminal block
TB-1 in the sterilizer power box. Use green/yellow wire for
European installations.
• 3-phase service requires a clearly marked disconnect with
lockout/tagout capability located near the sterilizer.
2.2.2 Sterilizer Final Check • Chamber must be leveled properly.
• Door must open and close smoothly.
• Door locked switches must be adjusted correctly.
• Chamber strainer must be in place.
• Rack and shelves and/or loading car operates correctly.
• Warranty labels properly applied.
2.2.3 Cycle Operation • Unit powers up correctly.
• Run Leak Test cycle–leak rate is to be less than 1.0 mm
Hg/minute (1.3 mbar/minute). Only on prevac sterilizers.
• Verify operation of a typical cycle (121° C[250°F] gravity).
2.2.4 Overall Size and Weight
(W x H x D)
• 16 x 16 x 26" Sterilizer: 660 x 1892 x 902 mm
(26 x 74.5 x 35.5")
• 20 x 20 x 38" Sterilizer: 762 x 1892 x 1152 mm
(30 x 74.5 x 45.375")