6-51
Sterilizer Operation Operator Manual 387366-389
Press the BASIC button, located on upper right of each extend screen
(see previous image) to return to the basic cycle parameters.
Table 6-1. Parameters with Changeable Value
Press applicable Parameter button and use keyboard display to enter new values.
Parameter Comments
1 Cycle Name Enter combination of letters and numbers to create
new name, then press ENTER. The new cycle name
appears on the Cycle Name button
2 Purge Time
(Excluding Isothermal,
USP660 and ATF 2)
Allowable range is 0 to 99 minutes and seconds.
Enter numeric value and press ENTER button. New
purge time is shown.
3 Pulses
(Prevac, Lab Waste,
ATF 1 and Decon
Only)
Allowable range is 0 to 99. Enter numeric value and
press ENTER button. New pulse value is shown.
4 Pulses
(Prevac, Lab Waste,
ATF 1 and Decon
Only)
Allowable range is 0 to 29.9 inHg. Enter numeric
value and press ENTER button. New pulse vacuum is
shown.
5 Pulse Charge
(Prevac, ATF1, ATF 2,
Lab Waste and Decon
Only)
Allowable range is 0 to 35.0 psig. Enter numeric
value and press ENTER button. New pulse charge is
shown.
6 Pulse Vac 1, 2 and 3
(Lab Waste and ATF 1
Only)
The vacuum levels can be varied from pulse 1, 2 and
3. Allowable range is 0 to 29.9 inHg. Enter numeric
value and press ENTER button. New pulse vacuum
value is shown.
7 Pulse Hold 1, 2 and 3
(Lab Waste, ATF 1 and
ATF 2 Only)
The vacuum hold time can be varied from pulse 1, 2
and 3. After vacuum level is reached, cycle controls
at the pressure for the time set. Allowable range is 0
to 59 seconds. Enter numeric value and press
ENTER button. New pulse hold value is shown.
8 Pulse Charge 1, 2
and 3
(Lab Waste, ATF 1 and
ATF 2 Only)
The pressure levels can be varied from pulse 1, 2
and 3. Allowable range is 0 to 35 psig. Enter numeric
value and press ENTER button. New pulse charge
value is shown.
9 Sterilize Time As applicable, press buttons for hours, minutes and
seconds to change sterilize time. Allowable range:
• Hours: 0 to 99
• Minutes: 0 to 59
• Seconds: 0 to 59
Enter numeric value and press ENTER button. New
sterilize time is shown.
10 Sterilize
Temperature
Allowable ranges:
• All Cycles except Isothermal:
100.0 to 141.0°C (212.0 to 285.0°F)
• Isothermal Cycles:
78.0 to 110°C (172.0 to 230.0°F)
Enter numeric value and press ENTER button. New
sterilize temperature is shown.