387366-389 Operator Manual Installation Verification
2.2.7 Utility Requirements
For Units Equipped With Optional
Electric Steam Generators
• Electric:
•Hot Water:
Pressure: 137.9 to 344.7 Kpa (20 to 50 psig)
Temperature: 60°C (140°F), maximum
Consumption: Peak 1gpm, per cycle 4 gal, Idle 1 gph
• Stainless-Steel Option Only:
Distilled, Reverse-Osmosis (RO), or Deionized Water with a
minimum specific resistivity of 1 m/cm.
• Cold Water:
Pressure: 206.8 to 344.7 Kpa (30 to 50 psig)
Temperature: 21°C (70°F), maximum
Consumption: Peak 6 gpm, per cycle 140 gal, Idle 10 gph
Control Box Power (Units with Steam Generator)
Voltage
(Vac)
# Phases Frequency Nominal
Current
Circuit
Protection*
120 1 50/60 Hz 9.0 A 20 A
Vacuum Pump Power (Units with Steam Generator)
Voltage
(Vac)
# Phases Frequency Nominal
Current
Circuit
Protection*
208
3 50/60 Hz
89.0 A 120 A
240 78.0 A 105 A
380 42.0 A 55 A
415 46.0 A 60 A
440 34.0 A 45 A
480 40.0 A 55 A
575 31.0 A 45 A
600 32.0 A 45 A
Water Ejector Power (Units with Steam Generator)
Voltage
(Vac)
# Phases Frequency Nominal
Current
Circuit
Protection*
208
3 50/60 Hz
83.2 A 110 A
240 72.2 A 95 A
380 38.0 A 50 A
415 42.0 A 55 A
440 31.0 A 40 A
480 37.0 A 50 A
575 28.0 A 40 A
600 29.0 A 40 A
* Circuit should be protected from exceeding the amp value shown, which is the maximum current rating.