387366-389 Operator Manual Operator Instructions
4.4 Manual Operation
of Door
Carefully review SECTION 1, SAFETY PRECAUTIONS, and the
following warnings and cautions before manually operating door(s).
WARNING - PERSONAL INJURY HAZARD: When closing the chamber door, keep hands and arms
out of the door opening and make sure opening is clear of obstructions.
WARNING - BURN HAZARD:
Steam may be released from the chamber when door is opened. Step back from the sterilizer each
time the door is opened to minimize contact with steam vapor.
Do not attempt to open the sterilizer door if a WATER IN CHAMBER ALARM condition exists. Call
a qualified service technician before attempting to use sterilizer further.
After manual exhaust, steam may remain inside the chamber. Always wear protective gloves,
apron, and a face shield when following emergency procedure to unload sterilizer. Stay as far
back from the chamber opening as possible when opening the door.
CAUTION - POSSIBLE EQUIPMENT DAMAGE HAZARD:
Gasket must be fully retracted prior to operating sterilizer door.
Make sure door opening is clear of any obstruction before closing the door(s).
Do not attempt to open sterilizer door during manual operation unless chamber is at 0 psig (0 bar).
Do not try to raise or lower door rapidly as fast operation may damage the manual door
mechanism.
Using hand pressure, pull up or push down on the door handle to
operate the door.
NOTE: Do not try to raise or lower door rapidly as fast operation may
damage the door drive mechanism.