6-21
Sterilizer Operation Operator Manual 387366-389
6.7 DART and
Bowie-Dick Cycles
Carefully review SECTION 1, SAFETY PRECAUTIONS, and the
Warnings and Cautions below before operating the sterilizer.
WARNING – BURN HAZARD:
Do not attempt to open the sterilizer door if a WATER IN CHAMBER ALARM condition exists. Call
a qualified service technician before attempting to use sterilizer further.
Sterilizer, rack/shelves, and loading car will be hot after cycle is run. Always wear protective
gloves and apron when removing a processed load. Protective gloves and apron must be worn
when reloading sterilizer following the previous cycle.
Steam may be released from the chamber when door is opened. Step back from the sterilizer
each time the door is opened to minimize contact with steam vapor.
After exhaust, steam may remain inside the chamber. Always wear protective gloves, apron, and
a face shield when following emergency procedure to unload sterilizer. Stay as far back from the
chamber opening as possible when opening the door.
When sterilizing liquids, to prevent personal injury or property damage resulting from bursting
bottles and hot fluid, you must observe the following procedures:
>>Use LIQUID cycle only; no other cycle is safe for processing liquids.
>>Use only vented closures; do not use screw caps or rubber stoppers with crimped seal.
>>Use only Type 1 borosilicate glass bottles; do not use ordinary glass bottles or any container
not designed for sterilization.
>>Do not allow hot bottles to be jolted; this can cause hot-bottle explosions. Do not move bottles
if any boiling or bubbling is present.
WARNING – EXPLOSION HAZARD: This sterilizer is not designed to process flammable
compounds.
WARNING – PERSONAL INJURY HAZARD: When closing the chamber door, keep hands and arms
out of the door opening and make sure opening is clear of obstructions.
CAUTION – POSSIBLE EQUIPMENT DAMAGE:
Gasket must be fully retracted prior to operating sterilizer door.
Ensure door opening is clear of any obstruction before closing the door(s).
Do not attempt to open sterilizer door during manual operation unless chamber is at 0 psig (0 bar).
Immediately wipe up saline solution spills on loading car, to prevent damage to stainless steel.
Sterilization of chloride-containing solutions (e.g., saline) can cause chamber corrosion and is
not recommended by the manufacturer. If, however, chloride-containing solutions must be
processed, clean the chamber after each use.
Failure to flush the integral steam generator (daily for carbon steel / every two weeks for stainless
steel) voids the manufacturer’s warranty and can lead to possible generator damage.