387366-389 Operator Manual Operator Instructions
The following emergency procedure should only be used in
instances where the sterilizer has lost either electrical or water
utilities, and a load is sealed in the chamber. This procedure
requires manually releasing the door seal by pressing on the door
and pushing the seal back into the groove.
1. Open front cabinet panel. Open manual exhaust valve to
exhaust remaining steam from the chamber (see F
IGURE 4-8).
Leave valve open during emergency procedure.
2. Using pressure tool provided, press on upper left hand and right
hand corners of chamber door as shown in F
IGURE 4-8. Door
should give inward slightly, indicating seal has been
compressed into groove.
3. Close front cabinet panel and pull down on door handle.
4. Once door is open, do not use sterilizer until unit has been
examined by a qualified service technician. Further use without
proper attention may damage sterilizer.
5. Close manual exhaust valve.
Figure 4-8. Emergency Door Operation
1. Check Chamber
Pressure Gauge
2. Open
Manual
Exhaust
Valve
3. Insert
Pressure
Tool
4. Press on Upper
Left-hand and Right-
hand Corners by
Pulling Pressure Tool
Away from Unit