4-5
Operator Instructions Operator Manual 387366-389
4.2.3 Operating Mode When sterilizer is placed in the Operating mode, the generated
printout lists the sterilizer type and manufacturer.
CONTROL ON
Day, Month XX, XXXX
XX:XX:XX AM
******************************
* STERIS SCIENTIFIC *
* LAB 110 / 250 SERIES *
* GRAVITY STERILIZER *
* MADE IN U.S.A. *
******************************
*******************************
Figure 4-6. Printout: Sterilizer Type
4.2.4 Cycle Start When a cycle is started, the generated printout lists name of cycle
started, time and date the cycle was started, the current cycle count
(number of cycles run since original start up of unit), operator name,
sterilizer ID number, default cycle number/type and the programmed
parameters for the cycle started.
NOTE: Cycle count value may be changed in the Supervisor Mode.
4.2.5 End-of-Cycle
Performance Summary
At the end of a cycle, the generated printout lists number of cycles
run that day, the maximum and minimum chamber temperatures
reached during the sterilize phase, processing times for key phases
and the total cycle time.
4.2.6 Alarm Condition When an alarm condition occurs, the generated printout (FIGURE 4-7)
lists the type of alarm and time, chamber temperature and chamber
pressure when it occurred.
NOTE: Refer to S
ECTION 7, ALARMS, for listing of possible alarm
conditions.
* ALARM
PRESSURE IN CHAMBER
F 10:07:23A 61.7C 34.0P
Figure 4-7. Printout: Alarm Condition
STERIS SCIENTIFIC
AMSCO
®
LAB SERIES
PREVAC STERILIZER