4-21
Operator Instructions Operator Manual 387366-389
4.9 Generator
Flush Alarms
4.9.1 List of Flush Alarms Generator Flush Alarms automatically abort the Generator Flush
Cycle. The alarms and their causes are listed in Table 7.1, ALARMS.
4.9.2 Setting “Too Long . . .”
Alarms
The following four “Too Long . . .” alarms can be set from the
Supervisor Mode screen by pressing the TOO LONG IN STEP
button. The allowable time range for each alarm is included below.
• TOO LONG TO PRESSURIZE GENERATOR
(1 minute to 120 minutes)v
• TOO LONG TO DEPRESSURIZE GENERATOR
(1 minute to 120 minutes)
• TOO LONG TO CLOSE GENERATOR DRAIN
(1 to 60 seconds)
• TOO LONG TO OPEN GENERATOR DRAIN
(1 to 60 seconds)
Refer to Table 6.1 for additional information.