6-78
387366-389 Operator Manual Sterilizer Operation
When a rate is enabled, it can be adjusted through use of the
associated Change Values screen. The Charge Rate allows pressure
rate to control from end of purge or end of last pulse to Sterilize phase.
Exhaust Rate allows pressure rate to control from end of Sterilize to
beginning of Vacuum Dry phase. Evacuate Rate allows pressure rate
to control from start of Vacuum Dry phase to end of Dry.
6.20.20 E-mail Alert System AMSCO
®
110LS and AMSCO
®
250LS Small Sterilizers include an
email alert system which can notify users, by email, of certain sterilizer
conditions and when preventive maintenance is due. Notification
topics are discussed later in this section of the manual.
To use the e-mail system, it must first be configured for use.
NOTE: E-mail setup is only required for first-time use and when
network and/or system-user changes must be made.
E-mail Setup
1. From Supervisor Mode Screen Page 5 of 5 (Screen 603), select
E-MAIL OPTIONS button.
NOTE: Discuss e-mail setup with your company’s IT Department
and/or Network Administrator before beginning setup procedure.
NOTE: First time the unit is powered up, it automatically tries
connecting to network.