6-4
387366-389 Operator Manual Sterilizer Operation
6.1.2 Sterilizer Equipped
with Loading Car
1. Open chamber door.
2. Verify loading car is securely fastened to transfer carriage.
3. Move transfer carriage forward until carriage latches with
chamber end frame (see F
IGURE 6-4).
Figure 6-4. Positioning Loaded Transfer Carriage
(Typical)
4. Ensure transfer carriage is latched to chamber end frame by
pulling carriage backward. If properly latched, carriage should
remain stationary.
5. Once carriage is latched in place, release loading car from
transfer carriage by lifting up carriage lock.
6. Carefully push loading car into sterilizer chamber. Ensure
loading car is positioned in back detent inside chamber.
7. Pull carriage latch knob to disengage transfer carriage from
chamber end frame. Move transfer carriage away from chamber.
8. Close chamber door. Sterilizer is now ready to run a processing
cycle. Refer to appropriate Cycle Operation Instructions
included in this section, for instructions on running the cycle.
Loading
Car
Transfer Carriage