6-52
387366-389 Operator Manual Sterilizer Operation
11 Over temp Allowable range is 0.0 to 40°C (32.0 to 104.0°F).
Enter numeric value and press ENTER button. New
sterilize temperature i s shown.
See SECTION 7, ALARMS, for explanation of over
temperature alarm.
12 Over drive Allowable range is 0.0 to 40°C (32.0 to 104.0°F).
Enter numeric value and press ENTER button. The
new sterilize temperature is shown.
See sterilize phase in cycle for an explanation of
control temperature.
13 Under temp. Allowable range is 0.0 to 40°C (32.0 to 104.0°F).
Enter numeric value and press ENTER button. The
new sterilize temperature is shown.
See SECTION 7, ALARMS, for an explanation of
under temperature alarm.
14 Print Interval Print interval is the time interval of status prints during
sterilize phase. Allowable range is 0 to 99 minutes.
Enter numeric value and press ENTER button. New
print interval is shown.
15 Resume/Restart RESUME resumes sterilize timer after it has been
stopped on an under temperature alarm.
RESTART restarts sterilize timer after is has been
stopped on an under temperature alarm.
Press either RESUME or RESTART button to resume
or restart. New resume or restart setting is shown.
16 Vacuum Dry
(Gravity, Prevac and
Lab Waste Only)
Vacuum dry setpoint is the vacuum level that dry
timer starts. Allowable range is 0.0 to 29.9 inHg (1.0 -
bar).
Enter desired numeric value and press ENTER
button.
New vacuum dry setting is shown.
17 Dry Time
(Gravity, Prevac and
Lab Waste Only)
As applicable, press buttons for hours, minutes and
seconds to change dry time. Allowable range:
• Hours: 0 to 99
• Minutes: 0 to 59
• Seconds: 0 to 59
Enter numeric value and press ENTER button. New
dry time is shown.
18 Exhaust Type
(Prevac, Lab Waste and
Decon Only)
The cycles have an option to run a fast or slow
exhaust after the sterilize phase. Press “Fast
Exhaust” section on Exhaust Type button to change
from Fast Exhaust to Slow Exhaust. Press “Slow
Exhaust” section on Exhaust Type button to change
from Slow Exhaust to Fast Exhaust.
19 Slow Exhaust First
Rate
(Liquid, Prevac,
Lab Waste, Isothermal
and Decon Only)
The slow exhaust rate is controlled at this value from
the end of sterilization to the slow exhaust first
transition point.
Table 6-1. Parameters with Changeable Value (Cont’d)
Press applicable Parameter button and use keyboard display to enter new values.
Parameter Comments