6-53
Sterilizer Operation Operator Manual 387366-389
20 Slow Exhaust First
Transition
(Liquid,
Prevac, Lab Waste,
Isothermal and Decon
Only)
Allowable range is 0 to 99.9 psig. Enter numeric
value and press ENTER button. New value is shown.
21 Slow Exhaust
Second Rate
(Liquid,
Prevac, Lab Waste,
Isothermal and Decon
Only)
The slow exhaust rate is controlled at this value from
the start of the slow exhaust first transition to the slow
exhaust second transition point. Allowable range is 0
to 99.9 psig. Enter numeric value and press ENTER
button. New value is shown.
22 Slow Exhaust
Second Transition
(Liquid, Prevac, Lab
Waste, Isothermal and
Decon Only)
Allowable range is 0 to 99.9 psig. Enter numeric
value and press ENTER button. New value is shown.
23 Vapor Removal
(Liquid Cycle Only)
Allowable range is 0 to 59. Enter numeric value and
press ENTER button. New value is shown.
24 Dry Pressure
Control
(Prevac, Lab
Waste, Gravity and Decon
Only)
Press Dry Pressure Control = Limit of System to
change the Dry Pressure Control to Vacuum Dry
Point. The vacuum system has the option to run to the
limit of system or control at the vacuum dry point
during the dry phase.
25 Charge Rate The charge rate is controlled at this value from the
start of the charge rate to the exhaust rate. Allowable
range is 0 to 99.9 psig. Enter numeric value and
press ENTER button. New value is shown.
26 Exhaust Rate The exhaust rate is controlled at this value from the
start of the exhaust rate to the evacuate rate.
Allowable range is 0 to 99.9 psig. Enter numeric
value and press ENTER button. New value is shown.
27 Evacuate Rate The evacuate rate is controlled at this value from the
start of the evacuate rate to the sterilize phase.
Allowable range is 0 to 99.9 psig. Enter numeric
value and press ENTER button. New value is shown.
28 Jacket
Temperature (Steam
Option Only)
Allowable range is 100 to 141.0°C. The new jacket
temperature is displayed.
29 Sterilize Control
(Load Probe Option)
Sets control to use drain or load probe temperature
to transition to sterilize phase. DRAIN option enters
sterilize phase when chamber temperature reaches
sterilize temperature. LOAD option enters sterilize
phase when chamber and load temperature reach
sterilize temperature. Selecting LOAD option also
enables Fo (See Fo cycle value).
30 F
0
(Load Probe Option)
Press Fo=0.0 to change the Fo value. The Charge
Phase or Sterilize Phase ends when the actual Fo
value is the Fo setting.
31 Interlock SeeS
ECTION 6.20.2, INTERLOCKS, for details.
Table 6-1. Parameters with Changeable Value (Cont’d)
Press applicable Parameter button and use keyboard display to enter new values.
Parameter Comments