6-54
387366-389 Operator Manual Sterilizer Operation
32 Air Detector
(Air Detector Option
Only)
The air detector value is the difference between the
chamber and air detector probes at the transition
between the vacuum to pressure pulses during the
Prevac cycle. Allowable range is between 0 to 99.9.
Enter numeric value and press ENTER button. New
value is shown.
33 Air Detector Pulses
(Air Detector Option
Only)
The air detector pulses is the number of vacuum
pulses during the Prevac cycle. Allowable range is
between 0 to 99. Enter numeric value and press
ENTER button. New value is shown.
34 Air Detector Active
(Air Detector Option
Only)
Turns on/off air detector feature during the Prevac
cycle. The air detector feature starts with a series of
vacuum pulses and then transitions to a series of
pressure pulses after it checks the air detector alarm
at the transition point (0 psig). Allowable range is ON
or OFF. Make selection and press ENTER button.
New value is shown.
35 Hold Time Allowable range is 0 to 999. Enter numeric value and
press ENTER button. New value is shown. (See
S
ECTION , CONTINUOUS CYCLES.)
36 Number of
Continuous Cycles
Allowable range is 0 to 9999. Enter numeric value
and press ENTER button. New value is shown. (See
S
ECTION , CONTINUOUS CYCLES.)
37 Charge Temperature
Rate (USP 660, ATF 1
and ATF 2 Only)
The rate controls the chamber temperature from the
end of active seal phase to the beginning of sterilize
phase. Allowable range is 0 to 99.9. Enter numeric
value and press ENTER button. New value is shown.
38 Exhaust Temperature
Rate (USP 660, ATF 1
and ATF 2 Only)
The rate controls the chamber temperature from the
end of sterilize phase to the beginning of deactivate
seal phase. Allowable range is 0 to 99.9. Enter
numeric value and press ENTER button. New value is
shown.
39 Preheat Temperature
(ATF 1 Only)
Allowable Range: 0 to 140°C
Enter numeric value and press ENTER button. New
Preheat Temperature displays.
40 Preheat Time
(ATF 1 Only)
Allowable Range: 0 to 120 minutes.
Enter numeric value and press ENTER button. New
Preheat Time displays.
41 Pulse Vacuum Hold
(ATF 1, ATF 2 and
Lab Waste Only)
Allowable Range: 0 to 99 minutes and
0 to 59 seconds.
Enter numeric value and press ENTER button. New
Pulse Vacuum Hold time displays.
42 Pulse Charge Hold
(ATF 1 and ATF 2 Only)
Pulse Charge Hold time can be varied from pulse 1,
2 and 3.
After pressure level is reached, cycle controls at
pressure for time set.
Allowable range: 0 to 59 minutes and 0 to 59
seconds. Enter numeric value and press ENTER
button. New Pulse Charge Hold Time displays.
Table 6-1. Parameters with Changeable Value (Cont’d)
Press applicable Parameter button and use keyboard display to enter new values.
Parameter Comments