6-14
387366-389 Operator Manual Sterilizer Operation
COMPLETE PHASE: Cycle is complete. Open door and unload
chamber. Once door is open, display returns to Main Operating Mode
Screen.
Example of Liquid Cycle tape.
==============================
========== LIQUID3 ===========
==============================
CYCLE START AT XX:XX:XXX
ON XX/XX/XX
CYCLE COUNT 0
LOGIN NAME: TOM
CYCLE NO. 2
STER TEMP = 121.0 C
CONTROL TEMP = 122.5 C
STER TIME = 0:45:00
V=inHg
- TIME T=C P=psig
------------------------------
C 11:48:24A 66.7 0.3V
C 11:49:24A 112.7 10.0P
S 11:49:43A 121.2 16.6P
S 11:51:43A 122.6 17.7P
S 11:53:43A 123.3 17.8P
S 11:55:43A 123.6 16.8P
S 11:57:43A 122.6 17.0P
S 11:59:43A 122.6 17.2P
S 12:01:43P 122.5 17.0P
S 12:03:43P 122.4 17.2P
S 12:05:43P 122.5 16.8P
S 12:07:43P 122.4 16.9P
S 12:11:43P 122.4 17.0P
S 12:13:43P 122.5 17.1P
S 12:15:43P 122.6 17.0P
S 12:17:43P 122.7 17.0P
S 12:19:43P 122.6 16.8P
E 12:19:44P 122.6 16.9P
Z 12:39:03P 89.9 5.1V
LOAD 080804
CHAMBER TEMP MAX=124.8 C
CHAMBER TEMP MIN=121.2 C
CONDITON = 1:19
STERILIZE = 30:01
EXHAUST = 21:42
TOTAL CYCLE = 42:01
==============================
======= READY TO UNLOAD ======
==============================