6-16
387366-389 Operator Manual Sterilizer Operation
NOTE: For the Isothermal Cycle, the following apply:
1) The Isothermal Cycle is designed for low temperature
processing (e.g., fractional sterilization, pasteurization) at 78 to
110°C (172 to 230°F).
2) The jacket is off during the Iso Mode.
3) Because the jacket is drained into the chamber, the sterilizer
door(s) must be closed before the Iso Mode button appears
(following figure, lower right), allowing unit to switch to Iso Mode.
4) If the jacket is less than 100°C (250°F), the jacket drain
phase is bypassed and Screen 4 is shown.
Run ISOTHERMAL CYCLE by pressing ISO MODE button on Cycles
screen.
Press wanted ISOTHERMAL button (values are for example only).
NOTE: Press LEFT ARROW button to cancel selected cycle and
return to Operating Mode Screen.
Press START CYCLE button to start the Isothermal Cycle.
NOTE: If CLOSE DOOR(S) button is shown, close door(s).
Start
Cycle