Section 4 Operation
4-76
2nd page
[Power Measure Method] F7: Sets the measurement method of Power Measure measurement , selecting Power
Meter or IF Level Meter.
[Power Control Bit Pattern] F8: Sets power control bits. (Initial value: Closed Loop)
Closed Loop: Performs power control of the TX level of the mobile station to
the desired value using the Closed Loop Power Control Bit.
Alternate : Sets power control bits to 0 or 1, alternately.
All 0 : Sets all power control bits to 0.
All 1 : Sets all power control bits to 1.
Auto: Performs the power control in real time operation.
This parameter setting affects only on the Power Meter screen.
[Closed Loop Power Control] F9:
Displays the closed loop power control menu. (Power Control Bit Pattern: Alternate)
[Closed Loop Up] F7: Increases the TX level of the mobile station by 1 dB.
[Closed Loop Down] F8: Decreases the TX level of the mobile station by 1 dB.
[User Define Level] F9: Sets the TX level of the mobile station to the User Define Level value on the
Setup TX Measure Parameter screen. In this time, the TX level is measured
using the IF Level Meter. So, the Reference Level must be set so that the TX
level and User Define Level can be measured.
[return] F12: Returns to the previous menu.
[Level Linearity Calibration] F10: Displays the level linearity calibration execution menu.
[Level Linearity Calibration] F7: Performs the level linearity calibration. (See the next page.) Calibration
progress window is displayed during calibration.
[Calibration Cancel] F8: Deletes the level linearity calibration data.
[return] F12: Returns to the previous menu.
[BS Output Level Cal.] F11: Performs the level calibration of the BS Output Level. During calibration, the output
signal becomes un-modulated.
[Back Screen] F12: Displays the previous screen.
*1: See paragraph 4.3.5 common parameter function.