4-75
4.3 CDMA Transmitter and Receiver Test --- TX and RX Tester Mode
For IF Level Meter
1st page
[Storage Mode] F9: Displays the Storage Mode Setting menu.
[Normal] F7: Sets normal mode.
[Average] F8: Sets averaging mode.
The Measure mode becomes Single, automatically.
[Average Count] F9: Sets an averaging count.
2 to 9999, Resolution: 1, Initial value: 10
An averaging count can also be set even if averaging mode is set off.
[Refresh Interval] F10: Sets the update time for averaged display.
Every: After each measurement
Once: After measurement with averaging
[return] F12: Returns to the previous menu.
[Calibration] F10: Displays the level calibration execution menu.
[Power Meter Calibration] F7: Performs the level calibration using an internal power meter.
Calibration progress window is displayed during calibration.
[Int. Osc. Calibration] F8: Performs the level calibration using an internal oscillator.
Calibration progress window is displayed during calibration.
[Calibration Cancel] F9: Deletes level calibration data.
[return] F12: Returns to the previous menu.
[Adjust Range] F11: Sets the measurement level range (RF power meter range and reference level) to a
state appropriate for the measurement signal. (See paragraph 4.3.1 (2))
[Back Screen] F12: Displays the previous screen.