4.3 CDMA Transmitter and Receiver Test --- TX and RX Tester Mode
Deviation function keys:
1st page
[Demod.] F7 Selects the demodulation function of FM (measurement unit: kHz) or øM (measure-
ment unit: rad).
Initial value: FM
[Relative On Off] F8 Displays the relative value with the reference value that is the measured level when
this key is pressed.
Initial value: Off
[Det Mode] F9 Selects the detection mode of:
1st page: (P-P)/2, +P, -P, RMS
2nd page: (P-P)/2 Hold, +P Hold, -P Hold
Initial value: (P-P)/2
[HPF] F10 Selects the HPF of 300 Hz, 50 Hz, or Off.
Initial value: Off
[LPF] F11 Selects the LPF of 3 kHz, 15 kHz, or Off.
Initial value: Off
[Back Screen] F12 Returns to the Setup Common Parameter screen.
2nd page
[Storage Mode] F9 Displays the Storage Mode menu for all the measured results on the screen.
[Normal] F7 Sets normal mode. (Initial value)
[Average] F8 Sets average mode.
[Average Count] F9 Sets number of Averaging processings.
2 Set value 9999
Initial value: 10
(In the average mode, the measurement is of single mode, which displays the
averaged results in each measurement, and stops measurement when the Aver-
age Count reached.)
Note that the Power Meter has not the average mode.
[return] F12 Returns to the Deviation menu.
[Back Screen] F12 Returns to the Setup Common Parameter screen.