Technical Information
175
APPENDIX
Accessories Skin temperature probe ....................................... 1
Oxygen sensor ...................................................... 2
Pneumoclean (Electrostatic air filter) ................. 1
Access port cover .................................................. 2
Piping connecting hose ........................................ 1
Dust cover .............................................................. 1
Operation Manual ................................................. 1
Temperature
Control mode Manual control/servo control (selectable)
Setting range Skin temperature (servo control): 34.0~37.5˚C
(override mode: 37.6~38.0˚C)
Incubator air temperature (manual control): 23.0~37.0˚C
(override mode: 37.1~39.0˚C)
Display range Skin temperature: 30.0~42.0˚C
Accuracy: 0.3˚C
Incubator air temperature: 20.0~42.0˚C
Accuracy: 0.3˚C
Heater output 0~100 %(indicated in 10 levels)
Warm-up time 60min at ambient temperature 25˚C
Alarms High temperature, set temperature, skin temperature probe
Humidity
Control mode Servo control
Setting range 40~95% Rh (in 1% increments)
Display range 15~99% Rh
Accuracy: 10%Rh
Continuous humidification without supplying water
8 hours
Maximum humidity 90%Rh (at ambient humidity 50% Rh, ambient temperature 25˚C, incubator air
temperature setting 37˚C)
Alarms Humidity sensor, low water level, no water, humidity chamber off, set humidity
Oxygen supply
Maximum oxygen concentration
65% O2 (at O2 flow rate 10L/min)
Environment
CO2 concentration in the hood When the air mixed with 4% CO2 is administered to a point 10cm above the cen-
ter of the mattress at 750mL/min and a stable state is achieved, the CO2 concen-
tration in the hood should not exceed 0.4%.