Maintenance Inspection
49
MAINTENANCE
3-3. Inspection Checklist
WARNING
If any trouble is detected in any inspection, indicate on the unit that it is out of order and seek repair
without fail.
Contact your local Atom representative for further information on repairs.
Judgment
YES / NO
YES / NO
YES / NO
YES / NO
YES / NO
YES / NO
YES / NO
YES / NO
YES / NO
YES / NO
YES / NO
YES / NO
YES / NO
YES / NO
YES / NO
YES / NO
YES / NO
YES / NO
YES / NO
YES / NO
YES / NO
YES / NO
YES / NO
YES / NO
YES / NO
No.
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
1
2
3
4
5
6
7
8
9
10
Item to check
Inspection before use (every time)
Are the main body and the hood free of breakage and other defects?
Is the main body fixed securely to the HL stand?
Are the access port covers free of breakage and other defects and attached securely?
Are the access port packings fixed securely to the access ports?
Are the tube introduction slit assembly packings free of breakage and attached securely?
Is the sensor module free of breakage and fixed securely?
Are the oxygen sensors free of any sign of damage, deterioration or leakage?
Are the admittance panel operating knobs held in place securely and do they operate reliably?
Are the access ports held in place securely and do they open and close reliably?
Does the tilting mechanism operate smoothly?
Does the power switch turn the power on and off reliably?
Is the fixing device of the control panel free of breakage and is the control panel fixed securely?
Is the filter clean without any blackish dirt?
Are all the caution/warning stickers affixed securely?
Is the Operation Manual readily accessible?
Quarterly inspection
Is the displayed incubator air temperature stable at 36.01˚C in manual control?
Is the displayed skin temperature stable at 36.00.5˚C in servo control?
Is the displayed RH stable at 9010%?
Does the oxygen concentration rise to 65% or higher at 10L/min?
Is the displayed oxygen concentration stable at 402% when the oxygen control function is on?
Is an SpO2 value displayed when an SpO2 sensor is connected?
Does the weight display show a correct numerical value when a 5kg weight (option) is weighed?
Do the “no water alarm” indicator and the “humidity chamber off alarm” indicator come on?
Is the fan free of breakage?
Does an audible alarm sound and does the power failure alarm indicator come on when the
power plug is removed from the power outlet?
Serial No. ( ) Date of Inspection ( . . ) Inspector ( )