Maintenance Inspection
48
3-2. Quarterly Inspection
Check the following items every three months.
Item to check
Incubator air
temperature control
Skin temperature
control
Humidity control
Oxygen concentra-
tion control
Pulse oximeter
Fan
Humidifier
Power failure alarm
Weight monitor
(The unit equipped
with the weight
monitor unit)
Procedure
Set the temperature to 36.0˚C in manual
control.
Place the skin temperature probe at about
10cm above the center of the mattress
surface in the incubator and set the tem-
perature to 36.0˚C in servo control.
Set RH to 90% with the incubator air tem-
perature set to 32.0˚C.
q
When the oxygen controller is not
used: Supply oxygen at 10L/min from
the oxygen supply port 1.
w
When the oxygen controller is used:
Connect oxygen to the oxygen supply
port 2 and, after 21% calibration, set the
oxygen concentration to 40%.
Connect a sensor to the main body and
attach the sensor to the inspector.
Check the exterior by visual inspection.
q
Install the cartridge tank without any
water in it.
w
Open the humidity chamber cover.
e
Install the humidity chamber with the
boiler cap off.
Turn the power on, and then remove the
power plug from the power outlet.
Place a 5kg weight (option) on the cen-
ter of the mattress and check the display.
Description
The displayed incubator air temperature
should be stable at 36.0
1˚C.
The displayed skin temperature should
be stable at 36.0
0.5˚C.
The displayed RH should be stable at
90
10%.
q
The oxygen concentration in the incu-
bator should rise to 65% or higher.
w
21% calibration should be completed
successfully, and the displayed oxygen
concentration should be stable at
40
2%.
95~100% should be displayed on the
%SpO2 display. The pulse rate should also
be displayed.
It should be neither broken nor de-
formed.
q
The “no water alarm” indicator should
come on.
w
The “humidity chamber off alarm” in-
dicator should come on.
e
The “humidity chamber off alarm” in-
dicator should come on.
The power failure alarm indicator should
come on and an audible alarm should
sound.
The displayed value should be in the
range of 5000
5g.