Technical Information
174
[10]Technical Information
10-1. Technical Data
Power requirements Customer-specified
쮿 for 200V region
Rating: AC230V; power consumption 600VA;
Frequency: 50/60Hz
Operating voltage range: AC230V10%
쮿 for 100V region
Rating: AC120V; power consumption 600VA;
Frequency: 60Hz
Operating voltage range: AC120V10%
Classification Type of protection: Class1 equipment
Degree of protection: Type BF applied part
Not for use in an air and flammable anesthetic gas mixture or an oxygen/nitrous
oxide and flammable anesthetic gas mixture
Mode of operation: Continuous operation (the HL stand is designed for inter-
mittent operation of 3 minutes per hour)
External display Type: Liquid crystal (TFT-LCD)
Size: 8.5 inches
Dot formation: 800 480 dots
Number of colors: 256 colors
Operating conditions Ambient temperature: 20~30˚C
Relative humidity: 30~75%
Atmospheric pressure: 70~106kPa
Wind velocity: 0.3m/s
Storage conditions Ambient temperature: 0~50˚C
Relative humidity: 30~75%
Atmospheric pressure: 70~106kPa
Maximum load capacity Dovetail rail: Approx. 20kg
I.V. pole (option): Approx. 10kg
Drawer (option): Approx. 3kg
❋ Total load capacity of the dovetail rail and the I.V. pole is approx. 30kg.
Dimensions Main body (with the HL stand):
68(W) 107(D) 138~178(H) cm (excluding the display)
Mattress surface 80~120cm in height
Mattress surface (with the weight monitor) 81.5~121.5cm in height
External display: 24.5(W) 6.5(D) 16(H) cm
Mattress: 65(W) 36.5(D) 2(T) cm
Weight Approx. 100kg (approx. 104kg if the unit includes the weight monitor)