Maintenance Inspection
42
3-1. Inspection before Use
The description in this section relates to a unit which is equipped with an oxygen controller and a weight
monitor. Please skip any section unrelated to your unit.
Check the following points every time you are going to use the unit.
CAUTION
If you should notice any strange smell, strange noise, overheating or strange vibration when the
power is applied to the main body in the inspection before use, stop using the unit immediately and
contact your local Atom representative.
3-1-1. Visual Inspection
Description
The main body and the hood should be neither broken nor deformed.
(Otherwise, the infant and/or the user may get injured due to breakage,
etc.)
The main body should be fixed securely to the HL stand with the connect-
ing bolts. (Otherwise, the main body may fall down.)
Each packing should be attached correctly to the access port. (Otherwise,
the air tight seal may be compromised.)
Each packing should be attached correctly to the hood. It should be intact.
(Otherwise, the air tight seal may be compromised.)
Each assembly should be attached correctly to the hood. It should be in-
tact. (Otherwise, the air tight seal may be compromised.)
It should be neither broken nor deformed. (Otherwise, unsatisfactory con-
trol may result from faulty detection by the sensors.)
Each knob should be held in place securely and should open and close the
admittance panel reliably. (Otherwise, the infant may fall out of the baby
compartment due to incomplete locking.)
It should be neither broken nor deformed. (Otherwise, the infant may fall
out of the baby compartment due to incomplete attachment.)
Each access port should be held in place securely and should open and
close reliably. (Otherwise, the infant may fall out of the baby compartment
due to incomplete locking.)
It should operate smoothly.
(Otherwise, it may not function.)
It should turn the power on and off reliably.
(Otherwise, it may not function.)
It should be fixed securely with the fixing device. (Otherwise, the control
panel may become unstable or fall when a switch is operated.)
Each caster should rotate smoothly. (Otherwise, the unit cannot be easily
relocated.)
It should be clean.
(Otherwise, air circulation may not be controlled properly.)
Item to check
Appearance
Connecting bolt
Snap-open access port
packing
Ventilator slot grommet
packing
Tube introduction slit
assembly
Sensor module
Admittance panel operating
knob
Baby guard
Snap-open access port
Tilting mechanism
Power switch
Control panel
Caster
Filter