Technical Information
180
Guidance and manufacturers declaration – electromagnetic immunity
The Incu i is intended for use in the electromagnetic environment specified below. The customer or the user
of the Incu i should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips,
short interrup-
tions and voltage
variations on
power supply
input lines
IEC61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-6
Electromagnetic environment guidance
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30 %.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment. If the user
of the Incu i requires continued operation during
power mains interruptions, it is recommended
that the Incu i be powered from an
uninterruptible power supply or a battery.
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
IEC 60601
test level
6kV contact
8kV air
2kV for power
supply lines
1kV for input/
output lines
1kV differential
mode
2kV common
mode
<5% UT
(>95% dip in UT)
for 0.5 cycle
40% UT
(60% dip in UT)
for 5 cycles
70% UT
(30% dip in UT)
for 25 cycles
<5% UT
(>95% dip in UT)
for 5 sec
3 A/m
Compliance
level
6kV contact
8kV air
2kV for power
supply lines
1kV for input/
output lines
1kV differential
mode
2kV common
mode
<5% UT
(>95% dip in UT)
for 0.5 cycle
40% UT
(60% dip in UT)
for 5 cycles
70% UT
(30% dip in UT)
for 25 cycles
<5% UT
(>95% dip in UT)
for 5 sec
3 A/m
NOTE UT is the a.c. mains voltage prior to application of the test level.